S-Blue vs T-Flex

Perfect Dental (Shenzhen Perfect Medical Instruments)

TR

S-Blue™

Perfect Dental (Shenzhen Perfect Medical Instruments)

Entry-Level Heat-Activated System

View protocol →

T-Flex™

Perfect Dental (Shenzhen Perfect Medical Instruments)

High-Purity NiTi Tuned for Curved Canals

View protocol →

Clinical Tip

Compare specifications above to choose the best system for your clinical case.

Specifications

Specifications comparison: S-Blue vs T-Flex
S-BlueT-Flex
ManufacturerPerfect Dental (Shenzhen Perfect Medical Instruments)Perfect Dental (Shenzhen Perfect Medical Instruments)
AlloyBlue Heat-Activated NiTiHigh-Purity Black NiTi (low carbon/oxygen)
Heat TreatmentBlue Heat-TreatedOther
MotionContinuous RotationContinuous Rotation
Files45
TaperVariableVariable
Cross-sectionCircular triangular (per brochure)Brochure shows a cross-section illustration but does not name the geometry
Speed425 RPM300 RPM
Torque2.5 Ncm1.5–2.5 Ncm
Use TypeSingle-useSingle-use

Best For

S-Blue

entry-level heat-treated rotary sequencegeneral posterior shaping with predictable motor settingspre-bendable files for moderate access challenges

T-Flex

curved canal preparationgeneral multi-file rotary shapingpractices wanting a balanced flexibility/cutting profile

Key Features

S-Blue

Four-file rotary sequence (B0 starter, B1, B2, B3) with circular triangle cross-section, uniform 425–500 rpm / 2.5 N·cm settings, and pre-bendable heat-activated NiTi (per brochure Design Concept and Specification table; brochure body says only "heat-activated material" — "Blue" is inferred from the product name)

T-Flex

Five-file rotary sequence (20/.10 opener, 15/.03, 20/.04, 25/.04, 25/.06) made from high-purity NiTi with low carbon and oxygen impurities, intended for curved anatomy and reduced canal transportation (per brochure)