TF4Traverse
ManufacturerPerfect Dental (Shenzhen Perfect Medical Instruments)Kerr Endodontics
AlloyNiTiNiTi
Heat TreatmentConventionalConventional
MotionContinuous RotationContinuous Rotation
Files43
TaperVariable8%, 6%
Cross-sectionNot specified by name in IFU textTriangular
Speed350 RPM500 RPM
Torque2 Ncm1.47 Ncm
Use TypeSingle-useSingle-use

Best For

TF4

rotary + reciprocating hybrid workflowclinicians comfortable switching motor modes mid-casemoderate-curvature anatomy

Traverse

rotary glide path creationpairing with TF Adaptive systemfaster glide path vs hand files alone

Key Features

TF4

Four-file hybrid sequence (T20 20/.07, T25 25/.07, T35 35/.06, T45 45/.05) where T25 25/.07 is run in reciprocating motion within an otherwise rotary workflow at 350 rpm / 2.0–3.0 N·cm (per IFU spec table and step-by-step)

Traverse

Rotary glide path system with orifice opener and two glide path file sizes — pre-sterilized, triangular cross-section, higher torsional resistance vs competitors (manufacturer comparison)