Skip to content

TF Adaptive™

Kerr Dental

TF Adaptive™

Adaptive motion automatically transitions between rotation and reciprocation based on canal stress

AlloyR-Phase NiTi (Twisted)File6

Specifications

Tap icons for term definitions
FileTipTaper
Speed
rpm
Torque
Ncm
SM10.204%AdaptiveAdaptive
SM20.256%AdaptiveAdaptive
SM30.354%AdaptiveAdaptive
ML10.258%AdaptiveAdaptive
ML20.356%AdaptiveAdaptive
ML30.504%AdaptiveAdaptive

Clinical Tips

Adaptive motion: rotation (600° CW) when unstressed, reciprocation (370° CW / 50° CCW) under load

R-Phase NiTi is twisted (not ground) — preserves grain structure, reduces microfractures

Files can be pre-curved using endo cold spray (cools to martensite phase)

SM1 doubles as a flexible glide path file for either sequence

TF Adaptive generates 58% less apical debris extrusion compared to WaveOne

Color-coded traffic light system: Green = Go, Yellow = Continue/Stop, Red = Stop

8:1 contra-angle handpiece required

Available in 23mm and 27mm lengths only

Warnings

  • REQUIRES Kerr Elements Motor or elements e-motion — do not use with standard rotary motors
  • Motor must be set to TF Adaptive program mode — never use continuous rotation
  • Contains nickel — do not use on patients with nickel allergy
  • File unwinding is a cautionary sign — reduce hand pressure or discard
  • Inspect for distortion or deformation before each use
  • Do not leave file at working length under load — advance and withdraw

Protocol

01Glide Path
02Shaping
03Finishing
For Educational Purposes Only

TF Adaptive™ is a trademark of Kerr Dental. This website is an independent educational resource and is not affiliated with, sponsored by, or endorsed by Kerr Dental.

Always refer to official manufacturer instructions before clinical use.

Checklist