ProDesign Logic®
Pre-bendable CM Wire with integrated glide path (.01 taper) and dual shaping series (.03 and .05) — file-specific speed/torque settings from 350 to 950 RPM
| File | Tip | Taper | Speed rpm | Torque Ncm |
|---|---|---|---|---|
| Glide Path | ||||
| 25/.01 | 0.25 | 1% | 350 | 1.0 |
| 30/.01 | 0.30 | 1% | 350 | 1.0 |
| Serie 03 | ||||
| 25/.03 | 0.25 | 3% | 600-950 | 2.0 |
| 30/.03 | 0.30 | 3% | 600-950 | 2.0 |
| 35/.03 | 0.35 | 3% | 600-950 | 2.0 |
| Serie 05 | ||||
| 25/.05 | 0.25 | 5% | 950 | 2.0 |
| 30/.05 | 0.30 | 5% | 950 | 4.0 |
| 35/.05 | 0.35 | 5% | 950 | 4.0 |
| 40/.05 | 0.40 | 5% | 950 | 4.0 |
| 25/.06 | 0.25 | 6% | 950 | 4.0 |
ProDesign Logic uses Controlled Memory (CM) NiTi wire (IFU describes the product as "Controlled Memory Rotary Nickel Titanium Files"). CM Wire files can be pre-bent to match canal curvature and will not spring back.
Speed and torque vary by file: .01 glide path at 350 RPM / 1 Ncm; Serie 03 at 600-950 RPM / 2 Ncm; Serie 05 at 950 RPM / 2-4 Ncm (25/.05 at 2 Ncm, 30-40/.05 at 4 Ncm per IFU table).
Available in 21mm and 25mm lengths only — 31mm lengths are not available (per IFU).
The system integrates glide path (.01 taper) and shaping (.03/.05 taper) into a single product family, eliminating the need for a separate glide path file system.
Single patient use only; sterilize before first use at 121°C for 20 minutes (per IFU).
Do not use sodium hypochlorite solution as a disinfectant for the files themselves (per IFU). NaOCl is used for canal irrigation, not instrument disinfection.
Obturation Techniques
Single cone, warm vertical, carrier-based & lateral compaction
Manufacturer page
Easy Equipamentos (MK Life) SystemsOther continuous rotation systems
Motion category
Continuous RotationComparisons