ProTaper Ultimate™
Slider-Shaper-Finishers concept with dual heat treatments: Gold for standard sequence, Blue for auxiliary finishers in large canals
AlloyGold/Blue Heat-Treated NiTiFile8
Specifications
Tap icons for term definitions
| File | Tip | Taper | Speed rpm | Torque Ncm |
|---|---|---|---|---|
| Slider | 0.15 | 2% (variable) | 300 | 1.5 |
| SX | 0.20 | 3% (variable) | 300 | 3.0 |
| Shaper | 0.20 | 4% (variable) | 300 | 3.0 |
| F1 | 0.20 | 7% (variable) | 300 | 2.0 |
| F2 | 0.25 | 8% (variable) | 300 | 2.0 |
| F3 | 0.30 | 9% (variable) | 300 | 2.0 |
| FX | 0.35 | 12% (variable) | 300 | 2.0 |
| FXL | 0.50 | 10% (variable) | 300 | 2.0 |
Clinical Tips
K-file free glide path: the Slider can be used directly in most cases
Only one Shaper needed for coronal 2/3 enlargement
Gold heat treatment on Slider/Shaper/F1-F3: +30% flexibility and fatigue resistance vs ProTaper Gold
Blue heat treatment on FX/FXL: dedicated for large straight canals
32% faster glide path and shaping compared to ProGlider + ProTaper Gold up to F2
Finishers will not unwind in the apical third
Use outward brushing motion away from furcation in irregular canals
Compatible with ProTaper Ultimate Conform Fit GP points and absorbent points
Warnings
- Single-use only — do not reprocess or reuse
- Contains nickel — do not use on patients with nickel allergy
- Discard if any unwinding, straightening, or stretching is observed
- Do not immerse in sodium hypochlorite solution
- FX and FXL are for large, straight canals only — do not use in curved canals
- Inspect cutting flutes upon removal from every canal
- Always reconfirm working length after shaping, especially in curved canals