Reference Hub

Endodontic Motor Settings

A cross-system summary of typical speed (RPM), torque (Ncm), and motion type for NiTi rotary and reciprocating files, with each row linking to the full system protocol.

Typical RPM by systemTorque guidanceContinuous · reciprocating · adaptiveAlways confirm the DFU

Motor settings are manufacturer-specific and depend on the file lot, market, and motor model. The table below aggregates the speed, torque, and motion values stored in each EndoGuide system module into a single reference summary. It is intended as an educational overview, not a directive: always confirm the exact program against the official DFU/IFU for the instrument you are using. Reciprocating systems run a preset auto-reverse program rather than a user-set RPM/torque, so those cells are shown as deferring to the DFU.

Speed, Torque & Motion by System

Values are aggregated from each system's per-file data and may combine several distinct settings across the file sequence. They are approximate educational guidance; the manufacturer DFU/IFU for the specific lot and motor is the authoritative source.

01ProTaper GoldDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

5.2 · 1.5 · 3.1

02ProTaper UltimateDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

4.0-5.2

03ProTaper NextDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

2-3

04WaveOne GoldDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

Per DFU

05RECIPROCVDW GmbH

Manufacturer

VDW GmbH

Torque (Ncm)

Per DFU

06Reciproc BlueVDW GmbH

Manufacturer

VDW GmbH

Torque (Ncm)

Per DFU

07HyFlex EDMColtene

Manufacturer

Coltene

Torque (Ncm)

2.5 · 1.8

08HyFlex CMColtene

Manufacturer

Coltene

Torque (Ncm)

2.5

09XP-endo ShaperFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

1.0

10TruNatomyDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

1.5

11EndoArt Actionİnci Dental (Turkey)

Manufacturer

İnci Dental (Turkey)

Torque (Ncm)

3.0 · 2.6 · 2.0

12EndoArt Touchİnci Dental (Turkey)

Manufacturer

İnci Dental (Turkey)

Torque (Ncm)

3.0 · 2.6 · 1.8

13EndoArt Smartİnci Dental (Turkey)

Manufacturer

İnci Dental (Turkey)

Torque (Ncm)

3.0 · 2.0 · 1.6

14EndoArt Expertİnci Dental (Turkey)

Manufacturer

İnci Dental (Turkey)

Torque (Ncm)

Per DFU

15VDW.ROTATEVDW GmbH

Manufacturer

VDW GmbH

Torque (Ncm)

1.3 · 2.1 · 2.3 · 2.0 · 3.5

16MtwoVDW GmbH

Manufacturer

VDW GmbH

Torque (Ncm)

1.2 · 1.3 · 2.1 · 2.3 · 1.6 · 2.0 · 3.0 · 1.0 · 1.7 · 0.3

17One ShapeMicro-Mega (COLTENE)

Manufacturer

Micro-Mega (COLTENE)

Torque (Ncm)

2.5

18One CurveMicro-Mega (COLTENE)

Manufacturer

Micro-Mega (COLTENE)

Torque (Ncm)

2.5

19TF AdaptiveKerr Dental

Manufacturer

Kerr Dental

Torque (Ncm)

Per motor

20ZenFlexKerr Dental

Manufacturer

Kerr Dental

Torque (Ncm)

1.00 · 3.0 · 2.0 · 3.5

21EdgeOne FireEdgeEndo

Manufacturer

EdgeEndo

Torque (Ncm)

Per DFU

22EdgeFile X7EdgeEndo

Manufacturer

EdgeEndo

Torque (Ncm)

300 g-cm

23R-MotionFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

Per DFU

24Vortex BlueDentsply Sirona

Manufacturer

Dentsply Sirona

Torque (Ncm)

75 · 104 · 132 · 195 · 290 · 368

252ShapeMicro-Mega (COLTENE)

Manufacturer

Micro-Mega (COLTENE)

Torque (Ncm)

2.0 · 2.5

26iRaCeFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

1.5

27EndoSequenceBrasseler USA

Manufacturer

Brasseler USA

Torque (Ncm)

1.8-2.3

28ProDesign LogicEasy Equipamentos (MK Life)

Manufacturer

Easy Equipamentos (MK Life)

Torque (Ncm)

1.0 · 2.0 · 4.0

29E-Flex GoldEighteeth

Manufacturer

Eighteeth

Torque (Ncm)

3.0 · 2.5

30BioRaCeFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

1.0

31iRaCe PlusFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

1.5

32RACE EVOFKG Dentaire

Manufacturer

FKG Dentaire

Torque (Ncm)

1.5

33One RECIMicro-Mega (COLTENE)

Manufacturer

Micro-Mega (COLTENE)

Torque (Ncm)

Per DFU

Manufacturer

EdgeEndo

Torque (Ncm)

2.94 Ncm (300 g-cm)

35EdgeEvolveEdgeEndo

Manufacturer

EdgeEndo

Torque (Ncm)

2.94 Ncm (300 g-cm)

36K3XFKerr Endodontics

Manufacturer

Kerr Endodontics

Torque (Ncm)

3.00 Ncm

37ESXBrasseler USA

Manufacturer

Brasseler USA

Torque (Ncm)

1.5-2.3

38MG3Perfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

3.0

39MG3 BluePerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

3.0 · 2.0

40Perfect TS4Perfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

2.0-3.0 · 1.0-1.5 · 1.0

41RCPerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

Per DFU

42RC GoldPerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

Per DFU

43RC BluePerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

Per DFU

44S-BluePerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

2.5

45T-FlexPerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

2.5 · 1.5 · 2.0

46T-ProPerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

Per DFU

47TF4Perfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

2.0-3.0

48TG6Perfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

3.0 · 1.5 · 2.5

49V-BluePerfect Dental (Shenzhen Perfect Medical Instruments)

Manufacturer

Perfect Dental (Shenzhen Perfect Medical Instruments)

Torque (Ncm)

Per DFU

How RPM & Torque Relate to File Behavior

Rotational speed (RPM) and the torque limit interact with the alloy and cross-section to influence cutting behavior and stress on the file. Lower torque limits generally cause the motor to auto-reverse or stop sooner when the file binds, which may reduce the risk of torsional overload in tight or curved canals.

Higher RPM can improve cutting efficiency but, in severe curvature, may increase the cyclic-fatigue load per unit time. Many manufacturers therefore pair a moderate RPM with a file-specific torque limit rather than a single global value — which is why torque can differ between files within one sequence.

Reciprocating systems do not expose a user RPM or torque in the same way: the motor runs a validated asymmetric angle program (a larger engaging stroke and a smaller releasing stroke) selected by the file's preset mode. For these systems, the safest practice is to select the manufacturer's named program and confirm it against the DFU.

Frequently Asked Questions

01

What RPM is typical for rotary NiTi files?

Answer

Many continuous-rotation NiTi systems are commonly run somewhere in the 250–600 RPM range, with some glide-path or orifice-opening files specified higher.

  • 01The exact value is system-specific, so the table above and the official DFU should guide the setting rather than a single global number.
02

How are reciprocating systems set?

Answer

Reciprocating systems typically run a preset auto-reverse program built into the motor rather than a user-entered RPM and torque.

  • 01You generally select the manufacturer's named mode (for example the file's dedicated reciprocation program) and confirm it against the DFU.
03

Why do recommended torque values differ between files in one system?

Answer

Torque limits are often tuned per file because tip size, taper, and the role of each instrument change how much stress it can tolerate.

  • 01A smaller apical file may carry a lower torque limit than a larger shaping file in the same sequence, which is why a single system can list a range of torque values.
04

Where do these numbers come from?

Answer

Each value is taken from the corresponding EndoGuide system module, which records the speed, torque, and motion published for that system.

  • 01Because manufacturers update instructions and figures vary by lot and market, these should be treated as an educational summary and verified against the current DFU/IFU.
05

Can I use the same settings on any endodontic motor?

Answer

Not necessarily.

  • 01Motor calibration, gear reduction in the handpiece, and how torque is measured can vary between devices.
  • 02It is generally advisable to enter the manufacturer's recommended program for your specific motor and to recalibrate the handpiece as instructed.