ZenFlex™
Violet oxide surface with 1mm flute diameter for minimally invasive shaping and excellent flexibility
AlloyHeat-Treated NiTi (Controlled Memory)File10
Specifications
Tap icons for term definitions
| File | Tip | Taper | Speed rpm | Torque Ncm |
|---|---|---|---|---|
| 20/.04 | 0.20 | 4% | 500 | 1.0 |
| 25/.04 | 0.25 | 4% | 500 | 1.0 |
| 30/.04 | 0.30 | 4% | 500 | 3.0 |
| 35/.04 | 0.35 | 4% | 500 | 3.0 |
| 40/.04 | 0.40 | 4% | 500 | 3.0 |
| 20/.06 | 0.20 | 6% | 500 | 2.0 |
| 25/.06 | 0.25 | 6% | 500 | 2.0 |
| 30/.06 | 0.30 | 6% | 500 | 3.5 |
| 35/.06 | 0.35 | 6% | 500 | 3.5 |
| 40/.06 | 0.40 | 6% | 500 | 3.5 |
Clinical Tips
All files use 500 RPM; torque varies by taper and size
Controlled-memory alloy provides excellent flexibility without shape memory spring-back
Violet titanium oxide surface layer is characteristic of the heat treatment
1mm flute diameter (20% smaller than competitors) for minimally invasive shaping
Can also be used in Adaptive Motion on Kerr elements e-motion motor
Available in 21mm, 25mm, and 31mm lengths for all sizes
Must be cleaned and sterilized before first use (non-sterile out of box)
Assorted packs available: .04 Asst (20-45) and .06 Asst (20-45)
Warnings
- Single-use only — do not reuse, risk of cross-contamination and breakage
- Contains nickel, brass, and titanium — do not use on patients with sensitivity
- Must be sterilized before first use: autoclave 132°C/4 min pre-vacuum or 121°C/30 min gravity
- Do NOT sterilize in chemical sterilization unit (causes corrosion)
- Maximum curvature: up to 90° — exercise additional caution at 30°+ curvature
- Inspect for deformation, breakage, fractures, or corrosion before use
- Use rubber dam to avoid accidental aspiration or ingestion
Protocol
01Glide Path
02Shaping
03Finishing
Powered by:Uzm. Dt. Levent Yüksel