EndoArt® Action
Variable taper design with Gold or Blue heat treatment options
| File | Tip | Taper | Speed rpm | Torque Ncm |
|---|---|---|---|---|
| Action Gold | ||||
| 0.19 | Variable | 300 | 3.0 | |
| 0.17 | Variable | 300 | 2.6 | |
| 0.20 | Variable | 300 | 2.6 | |
| 0.20 | Variable | 300 | 2.6 | |
| 0.25 | Variable | 300 | 2.6 | |
| 0.30 | Variable | 300 | 2.6 | |
| F4 | 0.40 | Variable | 300 | 2.6 |
| F5 | 0.50 | Variable | 300 | 2.6 |
| Action Blue | ||||
| SX | 0.19 | Variable | 300 | 2.6 |
| S1 | 0.17 | Variable | 300 | 2.0 |
| S2 | 0.20 | Variable | 300 | 2.0 |
| F1 | 0.20 | Variable | 300 | 2.0 |
| F2 | 0.25 | Variable | 300 | 2.0 |
| F3 | 0.30 | Variable | 300 | 2.0 |
| F4 | 0.40 | Variable | 300 | 2.0 |
| F5 | 0.50 | Variable | 300 | 2.0 |
Gold wire: x2 cyclic fatigue resistance vs conventional NiTi (per manufacturer product page)
Blue wire: x4 cyclic fatigue resistance vs conventional NiTi, allows pre-bending thanks to control memory (per manufacturer product page)
Blue files can be pre-bent before canal entry due to controlled memory
Speed: 300 RPM for all files. Torque differs by wire and file: Gold SX=3.0, Gold S1-F5=2.6; Blue SX=2.6, Blue S1-F5=2.0 Ncm
Laser-marked length indicators and tip/taper info on handle
Files may appear slightly curved after heat treatment - this is normal
Autoclavable for reuse (inspect before each use)
Obturation Techniques
Single cone, warm vertical, carrier-based & lateral compaction
From Inci Dental (Turkey) Systems
Manufacturer page
Inci Dental (Turkey) SystemsOther continuous rotation systems
Motion category
Continuous RotationComparisons
Alloy
Gold Heat TreatmentClinical guides